Table 11. Relative percent differences (RPD)
Table 10. Percent recoveries and relative percent differ-
for the sample duplicates.
ences established for the matrix spike duplicates.
Technology Developer
Techology developer
Sample
CRREL
HM 2000
CRREL
Visual
HM 2000
No./ID
(%)
(%)
Sample
% Recovery
% Recovery % Recovery
No./ID
(% RPD)*
(% RPD)
(% RPD)
1. Gasoline
SG-8 & 18
5.1
46
1. Gasoline
SG-10 & 19
0
67
Soil, SG-6 & 7
88
170
160
WG-11 & 14
0
140
(100 mg TPH/kg)
(4.5)
(0)
Soil, SG-20 & 20
91
98
120
2. Diesel
(500 mg TPH/kg)
(2.2)
`(34)
SDM-12 & 21
--
31
SDM-15 & 22
--
16
IF†
Water, WG-6 & 7
53
310
WDM-4 & 18
--
24
(0.48 mg TPH/L)
(28)
3. Motor oil
Water, WG-15 & 16
81
SL**
SL
M7 & M8
--
0
(24 mg TPH/L)
(15)
2. Diesel
Soil, SDM-6 & 7
--
180
310
design problems delayed and confounded several
(250 mg TPH/kg)
--
(10)
of the laboratory trials.
Soil, SDM-21 & 22
--
250
160
It has been stated that both the HM 2000 and
(1000 mg TPH/kg)
--
(38)
visual method of estimating TPH in environmental
Water, WDM-1 & 2
--
70
<5
matrices are capable of producing concentration
estimates within 10% of the concentration found
(0.98 mg TPH/L)
--
OR††
Water, WDM-16 & 17
--
170
by accepted methods of analysis, when the specific
(24 mg/kg TPH/L)
--
contaminant of concern is known (Hanby 1998).
However, this citation provided no experimental evi-
3. Motor oil
dence for this claim. This level of accuracy (10%),
Soil, M1 & M9
--
95
60
(1000 mg TPH/kg)
--
(12)
which is better than what has been reported else-
where on the basis of interlaboratory studies (Parr
* Relative percent difference.
†
Instrument failure.
et al. 1994), can be achieved by a single laboratory,
** Sample lost.
as demonstrated by the reference method (14 of 23
†† Over range, no dilution made.
cases had values within 10% for the analysis of
certified or spiked samples). In contrast, the HM
2000 only achieved this level of agreement for a certi-
occurs that quickly causes a very nauseating, chlorine-
fied or target value in 5 out of 34 cases (12%), and the
smelling gas to form. Therefore, cuvettes should only
visual method only achieved it in 2 out of 36 cases (5%).
be rinsed in an exhaust hood. A cleaning chamber, con-
In all cases where the HM 2000 or visual method gave
sisting of a washing station and reaction trap contain-
an average value within 10% of the certified or tar-
ing charcoal and sodium bisulfate, was fabricated by
geted concentration, the samples were spiked soils. Fig-
the technology developer to assist with this operation.
ure 3 shows the %D ranges of the values (33 visual, 31
However, in its current design, it failed to remove all
HM 2000, excluding less than or greater than values)
of the by-products, and additional rinses were neces-
found for both methods of analysis for all of the PE
sary once the cuvettes were removed from the cleaning
and spiked samples used in both the laboratory and field
chamber. Furthermore, we observed that these cuvettes
trials. This figure shows that roughly 60% of the visual
needed to be cleaned within a couple of hours of intro-
values for the spiked soils (12 of 21) and waters (7 of
ducing the catalyst, or the bottom window became per-
12), and the HM 2000 values for spiked soils (15 of
manently clouded. A combination of the time constraints
25), were within 50% of the certified or targeted val-
and need for an exhaust hood with the present cleaning
ues. However, only 1 out of 6 values (17%) attained by
chamber may limit the reuse of these cuvettes during
the HM 2000 for spiked water samples fell within this
field operations.
range. Clearly, neither method is capable of routinely
The instrumental problems experienced by the HM
achieving a 10% accuracy. With the exception of HM
2000 during the field exercise indicate that the present
2000 analysis of water matrices, perhaps they can
system is not robust, even when operated by the devel-
achieve a 50% level of accuracy, in the majority of
oper. Although not mentioned previously, instrumental
cases. It should be noted that, in comparison to the ref-
problems similar to those experienced in the field also
erence method for the analysis of field samples, this
plagued the laboratory trials. Intermittent instrument
9