Method B; however, two manufacturers did not
14 shows a schematic of the modified-cup appa-
specify the method, only that E96 was the stan-
ratus for water vapor permeance testing. This
dard used. Since agreement should not be ex-
method combines the features of both the ASTM
pected between results obtained by different meth-
dry and wet cup methods (Schwartz et al. 1989).
ods, it makes sense to specify a method and require
The modified test method creates the maximum
all membranes to be tested accordingly.
vapor pressure difference possible at the test tem-
perature of 84F (29C) that we used. This modi-
There are several critical aspects of the ASTM
E96 procedure that can adversely affect the per-
fied method eliminates the need to maintain a 50%
meance results. Toas (1989) reported on a 1985
RH environment on one side of the sample. That
round-robin test series using the E96 procedure.
saves a lot of effort in setting up and maintaining
He stated that in performing the ASTM E96 test,
a fixed relative humidity in the test chamber. Fig-
test operators must seal the test sample perfectly
ure 15 shows the modified-cup apparatus used for
to the test dish, carefully weigh the sealed speci-
permeance testing. The apparatus is held tightly
men with a balance having the proper sensitivity,
together by two aluminum plates bolted together.
and maintain the proper atmosphere for the test.
The assembly is placed in the test chamber and
Low permeance materials are difficult to measure,
periodically disassembled so that the water cup,
as the weight gain or loss is usually quite small
specimen, and desiccant cup portions can be
and the potential for error is much greater. Conse-
weighed. Then it is reassembled and the test con-
quently three material specimens are required, and
tinued.
a "dummy" should be used when testing materi-
We compared the results of the modified-cup
als with a permeance less than 0.05 perm. Statisti-
to a variation of the standard water test method
using a test temperature of 84F (rather than 73.4F
cally significant results require at least four rep-
[23C]) and an RH of 16% (rather than 50%). We
etitions of each membrane. These results usually
are determined using a least squares regression
also poked a hole through three of the membranes
analysis of the weight loss or gain, modified by
and re-tested these membranes using the water-
the dummy specimen as a function of time to ob-
cup method.
tain the water vapor transmission. We did not at-
tempt a statistical study as this would have been
Sample preparation
too costly. We decided instead to proof test an al-
Figure 16 shows a membrane sample ready for
ternative method and compare it to a variation of
testing. The membrane samples were adhered to
a standard method.
a stainless steel mounting plate using two tech-
niques. The first technique used pressure-sensi-
tive adhesives on five of the six membranes tested.
Procedures
We investigated a "Modified-Cup Method" as
A mounting jig was used to center these mem-
a result of our review and conversations with oth-
branes on their mask, and then the steel plate was
ers familiar with the E96 test procedures. Figure
pressed onto the membrane (after the release sheet
Table 5. Manufacturers' data for the membranes tested, sorted from lowest to highest permeance.
Permeance
(perm)
Thickness
ASTM test
(gr/hr ft2 in. Hg)
(g/s m2 Pa)
Manufacturer
Product tested
(mils)
method
1.71010
Protecto Wrap
M140A-R
60
E98
0.003
Method B
2.11010
Soprema
Sopralene Flam Antirock
170
E96
0.0036
2.9109
Royston
10AN Easy Pave ER
60
E96
0.05
Method B
4.6109
NEI
AC Bridge and Deck Seal
65
E96
0.08
Method B
5.7109
Polyguard
665 LT
65
E96
0.1
Method B
5.8108
W.R. Grace
Bituthene 5000
65
E96
1
16