APPENDIX B: DETERMINATION OF NITROAROMATICS, NITRAMINES,
AND NITRATE ESTERS BY SPE AND GC-ECD
1.0 SCOPE AND APPLICATION
2.3 GC-ECD determinations
Acetonitrile extracts are directly injected into the
This method is intended for the trace analysis
heated inlet of a gas chromatograph equipped with
an electron capture detector. The analytes are
chromatographic (GC) conditions for the detec-
resolved on a wide-bore fused-silica capillary
tion of the ppb levels using an electron capture
column coated with polydimethylsiloxane.
detector (ECD). All of these compounds are used
as explosives, are by-products of the manufac-
ture of explosives, or are the transformation
3.0 DEFINITIONS
to use by or under the supervision of analysts
Compound
Abbreviation
CAS* number
experienced in the use of GC-ECD, skilled in the
Octahydro-1,3,5,7-tetranitro-
interpretation of chromatograms, and experi-
1,3,5,7-tetrazocine
HMX
2691-41-0
enced in handling explosive materials. Also,
Hexahydro-1,3,5-trinitro-
each analyst must demonstrate the ability to gen-
1,3,5-triazine
RDX
121-82-4
1,3,5-Trinitrobenzene
TNB
99-35-4
erate acceptable results with this method.
1,3-Dinitrobenzene
DNB
99-65-0
2,4,6-trinitrophenylmethyl-
nitramine
Tetryl
479-45-8
2.0 SUMMARY OF METHOD
Nitrobenzene
NB
98-95-3
2,4,6-Trinitrotoluene
2,4,6-TNT
118-96-7
This method provides two optional proce-
4-Amino-2,6-dinitrotoluene
4-Am-DNT
1946-51-0
dures for preconcentration of low concentration
2-Amino-4,6-dinitrotoluene
2-Am-DNT
35572-78-2
(parts per trillion, or nanograms per liter) of cer-
2,4-Dinitrotoluene
2,4-DNT
121-14-2
tain explosives residues in surface or ground-
2,6-Dinitrotoluene
2,6-DNT
606-20-2
2-Nitrotoluene
2-NT
88-72-2
water. These options are cartridge solid-phase
3-Nitrotoluene
3-NT
99-08-1
extraction and disk solid-phase extraction.
4-Nitrotoluene
4-NT
99-99-0
3,5-Dinitroaniline
3,5-DNA
618-87-1
2.1 Cartridge solid-phase extraction
Nitroglycerine
NG
55-63-0
Pentaerythritoltetranitrate
PETN
78-11-5
For each sample, a Waters Sep-Pak Vac Pora-
pak RDX cartridge is placed on a solid-phase
*Chemical Abstract Service Registry Number
extraction vacuum manifold. The solid phase is
conditioned by passing acetonitrile through each
cartridge. Then reagent-grade water is pulled
through each cartridge to rinse away the acetoni-
4.1 Sample processing hardware
trile. Next, the analytes are extracted by pulling
Solvents, reagents, glassware, and other sam-
the aqueous samples through cartridges at low
ple processing hardware may yield artifacts
flow rate. Air is then pulled through each car-
and/or elevated baselines, causing misinterpre-
tridge to remove excess water. Finally, the car-
tation of the chromatograms. All of these materi-
tridges are eluted using acetonitrile.
als must be demonstrated to be free from inter-
ference.
2.2 Disk solid-phase extraction
For each sample, an Empore styrene-divinyl
4.2 Detector
benzene reverse-phase sulfonated resin (SDB-
In addition to nitrogenated organics, the ECD
RPS) disk is placed on a vacuum filter apparatus
will respond to other electrophores such as halo-
and preconditioned with acetonitrile. The aceto-
genated and oxygenated compounds. No speci-
nitrile is pulled through each disk, followed by re-
fic compounds have been identified as interfer-
agent-grade water, and then the water sample. Air
ences.
is then pulled through each disk to remove excess
water. Finally, the disks are eluted using acetoni-
4.3 Injector
trile.
The injection-port liner must be deactivated to
23