29
Vroblesky and Campbell (2001) for g/L and low mg/L concentrations. There-
ore, we believe that the reason the more volatile compounds were slightly higher
in the PDB samplers than in the standpipe was because they had volatilized from
the test solution. (The standpipe in these experiments was covered but not sealed,
and there were several inches of headspace.) Thus, the standpipe had lower
concentrations of these analytes at the time of sampling (i.e., the controls) while
the concentrations in the samplers reflected the previously higher concentrations
in the standpipe. The more volatile compounds had the largest differences.
Metals study
The variability between replicate samples was very low (<5% RSDs), except
for chromium concentrations in the PDB samplers where concentrations were
very low (Table 8).
Concentrations of metals in samples taken with the Discrete Interval
Sampler, HydraSleeve, Kabis Sampler, and PneumoBailer were generally not
significantly different from control values (Table 4). In the instances where there
were significant differences, the differences were less than 6%.
However, as with the pesticides, there were dramatic differences between the
metal concentrations in the PDB samplers and the controls. After 21 days' con-
act, three of the metals (As, Cd, Pb) were not detected in these samplers, and
concentrations of Cr were approximately 95% lower. Again, these results support
USGS guidance (Vroblesky 2001) that does not recommend using these devices
for inorganic ions. However, we should mention that other membranes, such as
the cellulose acetate (dialysis) membrane of the Multi-Level Sampler (available
from Margan Ltd.), can be used to sample metals and anions (Ronen et al. 1986,
1987a, 1987b; Magaritz et al. 1989, 1990; Ronen and Magaritz 1989; Kaplan et
al. 1991) or the Versapor membrane (acrylic copolymer coating over nylon
fabric) can be used for sampling colloids (Weisbrod et al. 1996).
Analytical results from the field study
Again, even though these samples had to be brought over 130 feet to the
surface and transferred to a sample vial, the variability was very low with the
Discrete Interval Sampler, PDB samplers, and HydraSleeve, with RSDs less than
4% (Table 9). Slightly greater variability was observed with the PneumoBailer
and the Kabis Sampler, with 15% and 8% RSD, respectively. The RSD for the
pumped control samples (using low-flow sampling) was less than 3%. With all
these devices, mean concentrations were 514% lower than the mean values for
the control samples.